REMOTE INTERFACE
99
This means that the magnitude of the DC offset is
determined by the output amplitude.
If the specified DC offset is out of range, the
maximum/minimum offset will be set instead. A
-222 error will be generated from the remote
terminal.
SOURce[1]:APPLy:SINusoid
Set
Outputs a sine wave when the command has
executed. Frequency, amplitude and offset can also
be set.
SOURce[1]:APPLy:SINusoid [<frequency>
[,<amplitude> [,<offset>] ]]
<frequency>
0.1Hz~25MHz*
<amplitude>
1mV~10Vpp (50)
<offset>
-5V ~ +5V (50)
*AFG-2005/2105 limited to 5MHz, AFG-2012/2112
limited to 12MHz.
SOURce1:APPL:SIN MAX, 3.0, -2.5
Outputs a 3Vpp sine wave at 25MHz (max
frequency) with a -2.5V offset.
SOURce[1]:APPLy:SQUare
Set
Outputs a square wave when the command has
executed. Frequency, amplitude and offset can also
be set. The duty cycle is fixed to 50%.
SOURce[1]:APPLy:SQUare [<frequency> [,<amplitude>
[,<offset>] ]]
<frequency>
0.1Hz ~ 25MHz*
<
amplitude
>
1mV~10V (50)